Introduction of SFY Series Moisture Analyzer

Chapter 1 Commissioning

2.1 Preparation before measurement

2.1.1 The moisture analyzer is placed on a fixed and horizontal workbench, loosen the gauze wire that protects the weighing part on the bottom of the instrument until the gauze head touches the desktop, lift the heating cylinder, and gently put the weighing bracket and scale Dial, adjust the two movable feet under the bottom plate of the instrument to make the instrument level.

2.1.2 Check that the AC power socket is intact, remove the power cord, and connect it to the instrument.

2.1.3 Turn on the power switch located at the back of the moisture analyzer. When the buzzer sounds, the instrument starts self-checking. After the 5 digits of the data display window display the numbers "99999-00000" in turn, it displays "SFY-2" . Then "0.000" is displayed, and the temperature display window displays the initial value.

2.1.4 The preheating method of the moisture analyzer must be preheated for 30 minutes before use. If the ambient temperature is low, it needs to be preheated for one hour. The data measured after preheating is true and valid.

2.1.5 Calibration of measurement accuracy. The instrument has been calibrated before leaving the factory. However, it is recommended that users use the oven method to calibrate the instrument when they use the instrument for the first time, or if they suspect that the results are in error and if the measurement object is a new variety. See Chapter 4 for details.

2.2 Parameter setting

2.2.1 Setting of discriminating time

Continue to press the "display" key (each time, the display content changes once), until the instrument "discrimination time" indicator lights, the instrument displays the initial value of 40 seconds, if you need to change, press the "↑" key to decrease Then press the "↓" key. After setting, press the "clear" key, the instrument returns to the current weight display state.

2.2.2 Heating temperature value setting

When you need to adjust the heating temperature value, there is a small door on the right side of the moisture meter. Press the door in the upper right corner of the small door and the door will pop up automatically. Open the door directly outwards, otherwise it will damage the structure of the small door switch.) Set the middle setting switch to the set position according to the adjustment instruction on the back of the small door. At this time, the setting indicator on the instrument panel lights. Then turn the setting knob to set a new temperature value (the heating temperature is set to 130 ° C when the instrument leaves the factory). Turn the switch back, and the normal indicator lights up. When measuring, if the actual temperature value does not reach the control temperature value, slowly turn the fine adjustment knob to calibrate. After adjustment, press the small door to close. (Temperature control has a lag. After rotating the fine-tuning knob, you should wait for a while to see if the displayed value reaches the set value. If not, turn the fine-tuning knob and wait until it is reached)

2.2.3 Connect printer and computer settings

After the instrument is powered on, the initial state is the printer-accessible mode. At this time, the optional printer can be connected. After measuring the moisture, press the "Print" key to print the moisture value. The instrument is not in the test moisture state. Press The "Print" key is invalid. If it is connected to a computer, press the "Display" key continuously until the "P 1" status is displayed, then press the "↓" key to make the instrument display "P 0". In this mode, the printer cannot be connected. When choosing to connect with a computer, the instrument will continuously send 6 bytes of ASCLL code data under the condition of measuring moisture. The first byte is the leading data "Y", and the last five bytes are high → low moisture value. The instrument must be turned off when connected to the indexer or computer, otherwise it will damage the instrument.

From: http: //